We developed a 5x5x5 mm3 crystal detector with an MMC readout. The detector was designed to achieve low energy threshold for direct detection of low mass dark matter. A pure CaF2 crystal was adopted as a target. This absorber crystal had a strong thermal contact to a metallic magnetic calorimeter (MMC) sensor via thin gold film evaporated on its surface. The MMC sensor and the gold film were directly bonded together using cold diffusion welding. We will present the result of the detector performance together with the detector model of the heat flow.