Gas multiplication process in high pressure proportional counters

24 May 2012, 19:21
<font color=green >La Biodola, Isola d'Elba, Italy</color=green><!-- ID_UTENTE=804 -->

<font color=green >La Biodola, Isola d'Elba, Italy</color=green><!-- ID_UTENTE=804 -->

<a href=http://www.elba4star.it>Hotel Hermitage</a> 57037 La Biodola Isola d'Elba (LI), Italy
Poster P6 - Gas Detectors Gas Detectors - Poster Session

Speaker

Dr Tadeusz Kowalski (AGH, University of Science and Technology, Cracow)

Description

Systematic measurements of gas amplification factors over the range 1 to 5 10$^5$, were performed for Ar + 20%CO$_2$ and Ar + 6.4% CO$_2$+ 2.5% N$_2$ mixtures as a function of the applied voltage between cathode and anode for the following working gas pressure: 0.05, 0.1, 0.3, 0.5 and 0.65 MPa. Full current characteristics, starting from recombination mode up to limited proportionality, have been measured. Single anode (radius of anode ra = 10, 15 and 25\, $\mu$m) cylindrical counters ( radius of cathode rk = 4, 5 and 6.5 mm) were used for the measurements. First Townsend coefficients were also determined. Obtained results will be presented.

Primary author

Dr Tadeusz Kowalski (AGH, University of Science and Technology, Cracow)

Co-author

Dr Stefan Koperny (AGH, University of Science and Technology, Cracow)