Description
The improvement of the timing performance is one of the main focus for several fields from big physics experiments to biomedical applications such as Time of Flight Positron Emission Tomography (ToF-PET). In this study we will present Single Photon Time Resolution (SPTR) and Coincidence Time Resolution (CTR) results of the recently introduced FBK NUV-HD Metal in Trench (MT) SiPM technology. Thanks to the addition of the optically insulating material inside the trenches, FBK NUV-HD-MT devices show an extremely low CrossTalk (CT). Moreover, the Photon Detection Efficiency (PDE) reaches the $\simeq65\%$ at $420nm$.
By using a femto-second laser with a wavelength of $390nm$, we have measured the SPTR for SPADs with different microcell sizes and different versions with a metal mask outside the active area (capacitive coupling). Moreover, a $1mm\times1mm$ and a $3mm\times3mm$ SiPM with $40\mu m$ cell size and M0 masking version have been tested. The CTR has been measured using a $4mm\times4mm$ SiPM to match the $3mm\times3mm\times5mm$ LYSO:Ce:Ca crystal. By using a high frequency readout electronics, we achieved a CTR of about $\simeq 80 ps$ FWHM and an outstanding SPTR of about $\simeq 19ps$ and $\simeq 30ps$ FWHM for the SPAD and $1mm\times1mm$ SiPM with $40\mu m$ M0 masking respectively.
This work opens the door to further investigations in order to study the role of the metal masking in the timing performance and to discuss about limitations and further improvements.
Field | Detectors and electronics |
---|