Speaker:
Giorgio Dho(Istituto Nazionale di Fisica Nucleare)
3
Report from WP3: Detector Simulation
Speaker:
Giulia D'Imperio(Istituto Nazionale di Fisica Nucleare)
4
Report from WP4: Detector Design and Construction
Speaker:
Giovanni Mazzitelli(Istituto Nazionale di Fisica Nucleare)
5
Report from WP5: Auxiliary Services
Speaker:
Andrea Messina(Istituto Nazionale di Fisica Nucleare)
6
Report from TC
Speaker:
Davide Fiorina(GSSI & INFN)
7
LIME Run Status
Speaker:
Stefano Piacentini(Istituto Nazionale di Fisica Nucleare)
LIME
Quest and EHD lens removed. Fusion and Schneider lens operational now.
Issue with the gas system: solved by power cycle of the system
Issue with the motor: solved by power cycle of the arduino
FIlter tests:
5 L/h fresh (tested also 2L/h and 1L/h)
50 L/h recirculation
Oxygen filters: duration ~ one week
Humidity: still working
Detector unstable behaviour: huge light yield (> 2x) with sparks and hot spots when oxygen concentration was as low as O(40 ppm). Forced to lower GEM voltage to 410V per GEM
Premixed bottles: arriving at LNGS tomorrow.
Next week plans: radon monitor. Premixed bottles?
8
Report from Editorial Board
Speakers:
Andrea Messina(Istituto Nazionale di Fisica Nucleare), Fabrizio Petrucci(Istituto Nazionale di Fisica Nucleare)
9
Report from Steering Committee
Speaker:
Fabrizio Petrucci(Istituto Nazionale di Fisica Nucleare)