The CUORE pulse tubes noise cancellation technique

23 Jul 2019, 17:45
1h 15m
Piazza Città di Lombardia (Milano)

Piazza Città di Lombardia

Milano

Piazza Città di Lombardia, 1, 20124 Milano MI
Poster Detector readout, signal processing, and related technologies Poster session

Speaker

Valentina Dompè (GSGC)

Description

The 1-ton scale CUORE detector is made of 988 TeO2 crystals operated as cryogenic bolometers at a working temperature of ~10 mK. In order to provide the necessary cooling power at 4K stage, a total of five Pulse Tubes refrigerators (PTs) are used. The PTs make the cryogenic system reliable and stable, but have the downside that mechanical vibrations at low frequecies (1.4 Hz and related harmonics) are injected into the experimental apparatus. We have developed an active noise cancellation technique in order to reduce such effect by taking advantage from the coherent interference between the pressure oscillations originated by different PTs. The technique that will be presented consists in controlling the relative phases of the pressure waves running inside the CUORE PTs lines, in order to achieve the lowest detector noise. By reducing the power of PTs harmonics by a factor up to 10^3-10^4, this technique allows to drastically suppress the overall noise RMS on the CUORE detector.