12–17 Oct 2015
Trieste - Italy
Europe/Rome timezone

Design and assembling of GEM detector sensitive volume for plasma radiation application.

13 Oct 2015, 16:15
Oceania (Trieste - Italy)

Oceania

Trieste - Italy

Congress Centre Stazione Marittima Molo Bersaglieri, 3 34123 Trieste Italy
Board: 29
Poster Production techniques Poster session & coffee break

Speaker

Dr Ewa Kowalska-Strzęciwilk (IPPLM)

Description

This abstract is devoted to design and assembling of GEM detectors for plasma radiation application. In this work we will report the results of mounting of two triple-GEM detectors with different dimensions 100x100mm2 and 200x20mm2. A description of the assembling procedures including the gluing/stretching techniques and selection of materials is given in this work. Particular attention was paid to the selection of materials that will work under special conditions, such as magnetic field, high temperature, neutron flux, electromagnetic and any type interference. Moreover, those materials can not disrupt detector’s operation, so to build it only certified materials should be used: checked under outgassing and ageing tests, radiation hardened, having appropriate electrical properties. The choice of materials was carried out so to have first of all the possibility of testing different materials (Ertacetal C and two types of fiberglass) for detector frames. There has also been the selection of assembly glue (Araldite) and the detector window material (thin aluminized Mylar foil).

Primary author

Co-author

Presentation materials

There are no materials yet.